Claims vs COA / AminoVault
AminoVault: claims vs COA
Website claim: 99% purity. “…Laboratory Compounds Made in the USA - Verifiable COAs with ≥99% Purity Guarantee Certified GMP - High Purity Guaranteed ISO/IEC 17…” Source (read 2026-09-25)
22
certificates read
9 / 14
vials outside ±3% of label
0
beyond ±10% (major)
5 / 22
below advertised purity
| Product | Label → measured | Purity | Main impurity | Lab | COA |
|---|---|---|---|---|---|
| BPC-157 | 10mg → 10.56mg Over label +5.6% | 98.8% 0.20 pts below 99% | 1.20% not itemised | ils-labs | Open |
| Epitalon | 10mg → 10.15mg On label +1.5% | 97.67% 1.33 pts below 99% | 2.33% not itemised | ils-labs | Open |
| Glutathione | 1500mg → 1581.15mg Over label +5.4% | 98.89% 0.11 pts below 99% | 1.11% not itemised | ils-labs | Open |
| Selank | 10mg → 10.45mg Over label +4.5% | 98.42% 0.58 pts below 99% | 1.58% not itemised | ils-labs | Open |
| COA-251469-QC251062-Klow.pdf | Blend, per-component content not listed | 98.75% 0.25 pts below 99% | 1.25% not itemised | — | Open |
| CJC-1295 No Dac | 10mg → 10.39mg Over label +3.9% | 99.21% Meets 99% | 0.79% not itemised | ils-labs | Open |
| Nad+ | 1000mg → 1048.07mg Over label +4.8% | 99.57% Meets 99% | 0.43% not itemised | ils-labs | Open |
| SS-31 | 50mg → 52.02mg Over label +4.0% | 99.63% Meets 99% | 0.37% not itemised | ils-labs | Open |
| Semaglutide GLP-1 | 10mg → 10.5mg Over label +5.0% | 99.65% Meets 99% | 0.35% not itemised | ils-labs | Open |
| Sermorelin | 5mg → 5.25mg Over label +5.0% | 99.62% Meets 99% | 0.38% not itemised | ils-labs | Open |
| Tesamorelin | 10mg → 10.33mg Over label +3.3% | 99.89% Meets 99% | 0.11% not itemised | ils-labs | Open |
| COA-251463-QC250891-TB-500.pdf | Not measured | 99.9% Meets 99% | 0.10% not itemised | — | Open |
| COA-251683-QC250898-Selank-Semax.pdf | Blend, per-component content not listed | 99.9% Meets 99% | 0.10% not itemised | — | Open |
| COA251470-QC251518-KPV.pdf | Not measured | 99.8% Meets 99% | 0.20% not itemised | — | Open |
| Ara-290 | 10mg → 10.22mg On label +2.2% | 99.58% Meets 99% | 0.42% not itemised | ils-labs | Open |
| BPC-157 / TB-500 | Blend, per-component content not listed | 99.86% Meets 99% | 0.14% not itemised | ils-labs | Open |
| GHK-Cu | 100mg → 102.66mg On label +2.7% | 99.42% Meets 99% | 0.58% not itemised | ils-labs | Open |
| Ipamorelin | Not measured | 99.62% Meets 99% | 0.38% not itemised | — | Open |
| KLOW | Blend, per-component content not listed | 99.8% Meets 99% | 0.20% not itemised | ils-labs | Open |
| MOTS-c | 10mg → 10.25mg On label +2.5% | 99.66% Meets 99% | 0.34% not itemised | ils-labs | Open |
| Semax | 10mg → 10.26mg On label +2.6% | 99.79% Meets 99% | 0.21% not itemised | ils-labs | Open |
| Wolverine | Blend, per-component content not listed | 99.86% Meets 99% | 0.14% not itemised | ils-labs | Open |
How this is checked
- Vial content. Measured net peptide content against the size on the label. Within ±3% is on label, 3–10% is off label, beyond ±10% is a major deviation (under- or overfilled). When a certificate weighs several vials of the batch, the vial farthest from the label is the one judged, and with several purity results the lowest counts. Blends are only checked when the certificate lists each component.
- Purity. HPLC purity on each certificate against the purity the supplier promises site-wide on its own website. Up to 1 point below the claim is amber, more than that is red. Only a direct promise counts (“99%+ purity guaranteed”); explanations of what purity means, one product's figure and page code are not treated as a claim. A certificate that reports only a bound (“>99%”) is judged only when the bound meets the claim, and a specification line is never read as a result.
- Whose certificate. A certificate is only read once our client-identity review has looked at it, and one issued to a different company is left out — whether the automatic check or a manual review found it. A certificate issued to the company that runs the supplier's own website counts as the supplier's. A file hosted on another supplier's website is not counted for this one. The same rules decide which certificates appear on supplier and product pages. Only peptide certificates are read: bacteriostatic water, acetic acid and other diluents are left out, and so are product data sheets, which state a specification rather than a batch result.
- Main impurity. Shown when the certificate itemises it: a named impurity (TFA, deletion sequence, oxidation…) or the largest secondary peak in the HPLC peak table. Most certificates report only a purity figure, so the rest of the sample is not itemised.
- Trust Score. These checks are 10 of the 25 Transparency points: vial content 7 (full credit within ±3%, half up to ±10%, none beyond, over- and underfilled alike), purity 3 (98% is the bar when a supplier advertises no figure). Fewer than 5 readable results gives half credit rather than a guess.
- These are the supplier's own published certificates, read automatically from the PDF or image. Janoshik results tables are read twice from the image, cell by cell, and a figure the two readings disagree on is left out rather than guessed. A certificate covers one batch on one date. We have not tested these batches ourselves.